<strong>Tsunami</strong><br />

To ensure the <strong>Tsunami</strong> laser can be tuned over its wavelength range with a<br />

fixed repetition rate, it is necessary to have a coarse cavity length adjust- -<br />

ment in addition to the fine length control provided by the PZT-controlled<br />

M4 mirror. Thus, mirror MI is mounted on a motorized micrometer and<br />

controlled through upldown buttons on the Model 3930. The MI control is<br />

a manual coarse control, while M4 is an automatic fine control. A bar<br />

graph indicator on the Model 3930 indicates the relative position of M4.<br />

The electronics module also provides controls and indicators for setting<br />

and viewing the photodiode signal level and the reference frequency. A<br />

built-in 10 Hz sweep generator can be used to create a 2 ns phase shift in<br />

the <strong>Tsunami</strong> photodiode signal. This facilitates locating pulses when<br />

trying to lock the LTC <strong>Tsunami</strong> to another laser.<br />

System Controls<br />

The following describes the displays and controls on the front and rear<br />

panels of the <strong>Tsunami</strong> Model 3930 LTC electronics module.<br />

Front Panel<br />

TIMING coarselfine control-sets timing between the reference source<br />

and the <strong>Tsunami</strong> laser. Both the coarse (outer knob) and fine (inner knob)<br />

controls have a 10-turn resolution. Range for the coarse control is approximately<br />

2 ns; it is 100 ps for the fine control. When the sweep generator<br />

(see below) is off, the coarselfine controls set the timing reference; when<br />

I<br />

-<br />

the sweep generator is on, the coarselfine controls are disabled.<br />

TIMING sweep generator onfoff push button-toggles the triangle wave<br />

sweep generator on and off (default = off). The sweep runs at about 10 Hz,<br />

and produces a linear timing change of about 2 ns. When the generator is<br />

off, the coarselfine controls set the timing reference; when on, the coarse1<br />

fine controls are disabled.<br />

LEVEL indicator-displays relative photodiode signal strength with<br />

full-scale = maximum signal. This is an uncalibrated reference indicator.<br />

f ,<br />

LEVEL FREQUENCY STATUS<br />

@ Spectra-Physics 'I<br />

Figure 8-2: Model 3930 Electronics Module Front Panel Controls and Indicators