<strong>Tsunami</strong><br />
With the exception of pump mirror PI and the GTI (ps configuration), all<br />
mirrors are captured and held in place by a screw-in holder. Unscrew the -<br />
holder and the mirror will come out with it. The optic is retained by a<br />
small O-ring and is removed by simply grabbing the exposed barrel and<br />
pulling it straight out of the holder. In most cases, you do not have to remove<br />
the optic from its holder to clean it.<br />
Each optical element has a v-shaped arrow on its barrel along with its part<br />
number. The arrow points to the coated surface that faces the intracavity<br />
beam. If you need to verify the location of the optic in the laser, refer to<br />
the tables located at the end of chapter 10, "Options and Accessories."<br />
If your laser becomes badly misaligned, refer to chapter 6, "Alignment,"<br />
for alignment procedures.<br />
The cleaning procedures that follow are placed in the order of importance<br />
that the optic(s) described contributes to laser output power loss.<br />
Brewster Windows<br />
Brewster-angle windows are incorporated at the input and output ends of<br />
<strong>Tsunami</strong> to seal the laser head, to allow it to be purged with nitrogen.<br />
From time-to-time the windows require cleaning.<br />
1. Close the pump laser shutter.<br />
2. Remove the windows.<br />
Remove the screws (2) holding each window in place, and remove the -<br />
windows by pressing them out from inside the cavity. Take care not to<br />
damage the window.<br />
3. Clean each surface.<br />
Use a folded tissue and hemostat according to step 3 of "General Procedures<br />
for Cleaning Optics."<br />
4. Install the windows.<br />
Referenced from the outside, the input window should face upward,<br />
and the output window downward.<br />
5. Install and tighten the mounting screws. Do not overtighten.<br />
6. Open the pump laser shutter.