<strong>Tsunami</strong><br />

Prism Translation Adjust<br />

Figure 6-16: Prl Prism Translation Adjust and Setscrew<br />

5. Set the position of Prl and Pr4 (Figure 6-10) to maintain the position<br />

of the fluorescent spots outside the cavity.<br />

The fluorescence from the beam might pass over the apex of Prl and<br />

Pr4, missing them, and pass through the AOM and out the laser. To<br />

ensure the alignment is performed correctly, move the prisms a bit<br />

further into the beam, temporarily, to make sure the fluorescence<br />

passes through them.<br />

a. Loosen the setscrew that locks prism Prl in place and move the<br />

prism up by turning its Prism Translation Adjust screw 3 to 4<br />

turns (Figure 6-16). Tighten the setscrew.<br />

b. Repeat step a for prism Pr4.<br />

c. Skip to step 7.<br />

6. Replace the GTI at MI (ps configuration).<br />

a. Turn off the Model 3955 electronics module.<br />

DANGER: HIGH VOLTAGE<br />

The pc board on the floor of the laser head contains high voltage. Turn<br />

off the Model 3955 electronics module to disable this board when<br />

changing the GTI.