Prism Translation Adjust<br />
Figure 6-16: Prl Prism Translation Adjust and Setscrew<br />
5. Set the position of Prl and Pr4 (Figure 6-10) to maintain the position<br />
of the fluorescent spots outside the cavity.<br />
The fluorescence from the beam might pass over the apex of Prl and<br />
Pr4, missing them, and pass through the AOM and out the laser. To<br />
ensure the alignment is performed correctly, move the prisms a bit<br />
further into the beam, temporarily, to make sure the fluorescence<br />
passes through them.<br />
a. Loosen the setscrew that locks prism Prl in place and move the<br />
prism up by turning its Prism Translation Adjust screw 3 to 4<br />
turns (Figure 6-16). Tighten the setscrew.<br />
b. Repeat step a for prism Pr4.<br />
c. Skip to step 7.<br />
6. Replace the GTI at MI (ps configuration).<br />
a. Turn off the Model 3955 electronics module.<br />
DANGER: HIGH VOLTAGE<br />
The pc board on the floor of the laser head contains high voltage. Turn<br />
off the Model 3955 electronics module to disable this board when<br />
changing the GTI.