Installation<br />
3. Connect the purge line from the Model 39 10 purge unit to the<br />
<strong>Tsunami</strong> laser head.<br />
Use the 3.5 m PTFE purge line provided to connect the Model 3910 to<br />
the laser head. It comes with a Clip-LokTM connector on one end and<br />
a quick-disconnect connector on the other. The purge line is attached<br />
by pressing the hose connectors onto their mating connectors until<br />
they snap in place.<br />
a. Attach the Clip-LokTM connector to the output of the Model 3910.<br />
b. Attach the quick-disconnect connector to the purge input on the<br />
<strong>Tsunami</strong> laser head input bezel.<br />
WA RNlNG<br />
Do not connect the Model 3910 to a boil-off tank or other non-lab-grade<br />
nitrogen gas source, nor to any source with a pressure greater than 67<br />
kPa (10 psi) or damage to the filters will result.<br />
4. Connect the gas supply line from the Model 3910 purge unit to the<br />
nitrogen supply. Make sure you use dry, oil-free, Electronic Grade 5<br />
(or better) nitrogen (99.999% pure) to prevent contamination of the<br />
Model 3910 or your laser optics.<br />
Use the 2 m, in. i.d. PTFE gas supply line provided to connect the<br />
Model 3910 to the nitrogen supply. It comes with a Clip-LokTM connector<br />
on both ends.<br />
a. Attach one end of the gas supply line to the INPUT FROM GAS<br />
CYLINDER connector on the Model 39 10.<br />
b. Attach the other end of the gas supply line to your nitrogen supply.<br />
If your nitrogen supply does not accept a Clip-LokTM connector,<br />
cut off the existing connector and install your own. A Clip-LokTM<br />
connector for in. i.d. PTFE tubing is also included in the accessory<br />
kit.<br />
To attach barbed connectors to PTFE tubing, heat the tubing with<br />
a flameless heat source (heat gun), then quickly slide the tubing<br />
onto the barbed fitting while the tubing is still warm. Do not<br />
move the connection until it has cooled. Install the line and check<br />
for leaks.<br />
This completes the installation of the Model 3910 purge unit.<br />
Clip-Lokm is a registered trademark of Anarak, Inc.