I 1 1 ill : I I I iil 1<br />

Operation<br />

-.<br />

I<br />

Operating with Lok-to-Clock"<br />

Refer to chapter 8 for installation and operating instructions for the<br />

optional Lok-to-Clock system.<br />

System Shut-down<br />

1. Turn off the pump laser according to its user's manual.<br />

2. Turn off the chiller.<br />

To prevent condensation on the surface of the rod, do not run the<br />

chiller when the laser is off.<br />

3. Turn off the nitrogen supply at the tank, then close the regulator valve<br />

on the Model 39 10 purge unit.<br />

Do not turn off the Model 3955 or the optional Model 3930 if you plan on<br />

using the laser within the next week or so. The Model 3930 takes about 2<br />

hours to warm up properly for stable 80 MHz operation.<br />

This completes the system shut-down procedure.