<strong>Tsunami</strong><br />

Selecting, Installing, and Aligning<br />

the Birefringent Filter (ps Systems Only)<br />

There are two birefiingent filter (bi-fi) sets that cover the <strong>Tsunami</strong> ps wavelength<br />

range. The "A" plate bi-fi has a yellow dot on the filter ring, the "B"<br />

plate bi-fi has a white dot. Use the "A" plate with the blue, standard, long,<br />

or extra-long optic sets. Use the "B" plate with the midrange optic set.<br />

CAUTION<br />

The alignment of each filter stack is set at the factory and does not require<br />

adjustment. Never try to disassemble the stack.<br />

Once you have chosen a wavelength at which to operate, select a filter and<br />

follow the procedure below to install and align it. Use Table 6-2 to select<br />

your filter.<br />

A monochromator is required for this procedure.<br />

1. Close the pump laser shutter.<br />

2. Set the bi-fi micrometer to it maximum clockwise position.<br />

3. Install the appropriate filter in the holder.<br />

a. Place the filter in the holder.<br />

b. Rotate the filter until the pair of setscrew holes on the outer edge<br />

of its ring (Figure 6-3) is near the position listed in Table 6-2 for<br />

your wavelength. Note: the "12 o'clock" position is the highest<br />

point and is nearest you as seen from the output bezel. This gets<br />

you close.<br />

c. Open the pump laser shutter.<br />

d. While watching the monochromator throughput from the<br />

<strong>Tsunami</strong>, rotate the filter slightly by hand until you obtain the<br />

lower wavelength listed in Table 6-2.<br />

e. Close the pump laser shutter.<br />

Table 6-2: Bi-Fi Alignment for the Nominal 2 ps Systems<br />

Optic Set<br />

Plate Type1<br />

Dot Color<br />

Setscrew<br />

Position (o'clock)<br />

Lowest<br />

Wavelength<br />

Highest<br />

Wavelength<br />

*<br />

Blue (8)<br />

Standard (S)<br />

ANellow<br />

ANellow<br />

1 :00 to 2:00<br />

12:OO to 1 :00<br />

688 nm<br />

715 nm<br />

790 nm<br />

850 nm<br />

Mid-range (M)<br />

BANhite<br />

1 :00 to 2:00<br />

775 nm<br />

900 nm<br />

Long (L)<br />

X-long (X)<br />

ANellow<br />

ANellow<br />

1 :00 to 2:00<br />

12:OO<br />

835 nm<br />

965 nm<br />

1000 nm<br />

1080 nm