<strong>Tsunami</strong><br />
Selecting, Installing, and Aligning<br />
the Birefringent Filter (ps Systems Only)<br />
There are two birefiingent filter (bi-fi) sets that cover the <strong>Tsunami</strong> ps wavelength<br />
range. The "A" plate bi-fi has a yellow dot on the filter ring, the "B"<br />
plate bi-fi has a white dot. Use the "A" plate with the blue, standard, long,<br />
or extra-long optic sets. Use the "B" plate with the midrange optic set.<br />
CAUTION<br />
The alignment of each filter stack is set at the factory and does not require<br />
adjustment. Never try to disassemble the stack.<br />
Once you have chosen a wavelength at which to operate, select a filter and<br />
follow the procedure below to install and align it. Use Table 6-2 to select<br />
your filter.<br />
A monochromator is required for this procedure.<br />
1. Close the pump laser shutter.<br />
2. Set the bi-fi micrometer to it maximum clockwise position.<br />
3. Install the appropriate filter in the holder.<br />
a. Place the filter in the holder.<br />
b. Rotate the filter until the pair of setscrew holes on the outer edge<br />
of its ring (Figure 6-3) is near the position listed in Table 6-2 for<br />
your wavelength. Note: the "12 o'clock" position is the highest<br />
point and is nearest you as seen from the output bezel. This gets<br />
you close.<br />
c. Open the pump laser shutter.<br />
d. While watching the monochromator throughput from the<br />
<strong>Tsunami</strong>, rotate the filter slightly by hand until you obtain the<br />
lower wavelength listed in Table 6-2.<br />
e. Close the pump laser shutter.<br />
Table 6-2: Bi-Fi Alignment for the Nominal 2 ps Systems<br />
Optic Set<br />
Plate Type1<br />
Dot Color<br />
Setscrew<br />
Position (o'clock)<br />
Lowest<br />
Wavelength<br />
Highest<br />
Wavelength<br />
*<br />
Blue (8)<br />
Standard (S)<br />
ANellow<br />
ANellow<br />
1 :00 to 2:00<br />
12:OO to 1 :00<br />
688 nm<br />
715 nm<br />
790 nm<br />
850 nm<br />
Mid-range (M)<br />
BANhite<br />
1 :00 to 2:00<br />
775 nm<br />
900 nm<br />
Long (L)<br />
X-long (X)<br />
ANellow<br />
ANellow<br />
1 :00 to 2:00<br />
12:OO<br />
835 nm<br />
965 nm<br />
1000 nm<br />
1080 nm