~lpl~<br />

f \<br />

<strong>Tsunami</strong><br />

Opto-Mechanical Controls<br />

ps Configuration<br />

The following describes opto-mechanical controls specific to <strong>Tsunami</strong><br />

lasers configured for ps operation. Refer to Figure 4-2 and Figure 4-3.<br />

MI Gires-Tournois interferometer (GT1)-used to compensate for dispersion<br />

and to adjust for the shortest pulse. The GTI replaces the HR used<br />

in the fs configuration. Its vertical and horizontal adjustments allow you to<br />

align the laser cavity and to optimize output power and mode quality.<br />

These controls are available when the laser head cover is in place.<br />

Birefringent filter (bi-fi) wavelength selector-used to select the laser<br />

wavelength. The micrometer control is available when the laser head cover<br />

is in place.<br />

Model 3955 Electronics Module Controls<br />

Front Panel<br />

The following is a description of the displays and controls on the front<br />

panel of the Model 3955. Refer to Figure 4-5.<br />

P HOT0 Dl 0 D E indicator--displays relative photodiode signal strength with<br />

full-scale = maximum signal. This is an uncalibrated reference indicator.<br />

STATUS: PULSING mode locker indicator-when lit, it indicates the<br />

system is mode locked. When off, it indicates no pulses are detected.<br />

STATUS: ENABLE mode locker indicator-when<br />

locking is enabled.<br />

lit, it indicates mode<br />

STATUS mode locker enableldisable button-turns on or off the rf power<br />

to the AOM. Note: even when rf power is off, power to the heater is still<br />

applied.<br />

PHASE display-indicates the relative position of the fine PHASE control.<br />

This is an uncalibrated reference indicator.<br />

L<br />

L PHOTODIODE STATUS PHASE POWER J<br />

@ Sp~ctrs-ChysBes j ;<br />

Figure 4-5: Model 3955 Electronics Module Front Panel Controls and Indicators 11 -,