Tsunami - Beckman Institute Laser Resource Center
Tsunami - Beckman Institute Laser Resource Center
Tsunami - Beckman Institute Laser Resource Center
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Tsunami</strong><br />
Removing the Purge Line<br />
from the <strong>Laser</strong> Head and Model 3910<br />
Installing the <strong>Tsunami</strong> Model 3955<br />
Electronics Module<br />
To release the purge line from the laser head, press inward on the metal<br />
locking flange on the quick-disconnect, while gently pulling on the hose.<br />
To release the purge line from the Model 3910 purge unit, press inward<br />
on the wire clip-spring while gently pulling on the hose.<br />
1. Verify the power switch on the rear panel is off.<br />
2. Verify the voltage select switch on the rear panel is set to match the ac<br />
line voltage of your facility outlet.<br />
The white indicator for the voltage setting is located next to the power<br />
cord connector. Voltage choices are 100, 120,220, and 240 V,,. If the<br />
voltage setting does not match your facility outlet voltage, refer to<br />
appendix C, "Setting the Line Voltage Switch."<br />
WARNING<br />
An improper voltage selection will damage the electronics.<br />
3. Place the Model 3955 electronics module in a convenient location<br />
within 3 m of the laser head.<br />
4. Attach one end of the signal cable (provided in the accessory kit) to<br />
the male D-sub TO 3955 connector on the input bezel of the laser head<br />
(Figure 4-3), and the other end to the TO HEAD connector on the rear<br />
panel of the Model 3955 (Figure 4-6).<br />
5. Connect one of the BNC cables (provided in the accessory kit) to the<br />
ML PD photodiode connector on the input bezel of the laser head<br />
(Figure 4-3) and the other end to the PHOTODIODE connector on the<br />
rear of the Model 3955.<br />
CA UTlON<br />
Do not connect the PHOTODIODE BNC cable to the LTC PD connector!<br />
-<br />
WARNING<br />
Be careful when connecting the photodiode and mode locker cables.<br />
Connecting the Model 3955 MODE LOCKER rf output cable to the<br />
<strong>Tsunami</strong> ML PD mode locker photodiode connector will cause permanent<br />
damage to the photodiode pc board.