<strong>Tsunami</strong><br />

b. Move the stage by hand and set it so the vertical edge facing the<br />

rod is placed at the third reference mark from the end of the slide.<br />

There are reference marks scribed on the vertical surface of the<br />

nonrnoveable dovetail. They are very fine and can be hard to see.<br />

c. Tighten the setscrew.<br />

10. Set the initial position for cavity focusing mirror M2 by adjusting the<br />

M2 micrometer for "7" on the gradation scale (Figure 6-5).<br />

1 1. Align pump beam steering mirror PI.<br />

a. Venfy the pump laser is set for minimum current.<br />

b. Open the pump laser and <strong>Tsunami</strong> shutters.<br />

c. <strong>Center</strong> the pump beam on Pl (Figure 6-6) using the vertical and<br />

horizontal adjustments on the routing mirrors.<br />

d. Adjust PI vertically and horizontally to center the pump beam<br />

on P2.<br />

12. Align pump beam focusing mirror P2.<br />

Rod Translation<br />

Control<br />

M2 Translation Pump Beam ~'riz. v&.<br />

, \<br />

Horiz. Vert.<br />

Control Dump M2 Cavity Focus Mirror M3 Cavity Focus Mirror<br />

Figure 6-5: Cavity Focus Mirrors Mq and Mg and the <strong>Laser</strong> Rod<br />