Tsunami - Beckman Institute Laser Resource Center
Tsunami - Beckman Institute Laser Resource Center
Tsunami - Beckman Institute Laser Resource Center
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Service and Repair<br />
Symptom: <strong>Laser</strong> is mode locked but has noisy >2 ps pulses with side lobes or baseline noise<br />
(ps configuration only)<br />
Possible Causes<br />
Poor cavity mode.<br />
Contact your Spectra-Physics <strong>Laser</strong>s field service representative. Do not at-<br />
GT' might be misaligned' tempt disassembly You will void your warranty.<br />
Pump power is too high.<br />
Possible oxygen or<br />
moisture contamination<br />
from the purge<br />
Corrective Action<br />
Verify M2 and M3 are properly aligned. Refer to chapter 6, "Alignment: Cavity<br />
Alignment."<br />
Reduce pump power until the side lobes or the baseline noise disappears.<br />
Use only dry, oil-free, Electronic Grade 5 nitrogen gas from a cylinder. Avoid<br />
boil-off tanks which might contain oxygen or water vapor. Check the color of<br />
the desiccants in the molecular-sieve assembly in the Model 391 0 and make<br />
sure they are not pink. If they are, replace the assembly (refer to chapter 9).<br />
Symptom: Pulses are broad with low power<br />
Possible Causes<br />
Fine PHASE adjust is not<br />
set properly.<br />
' and photodiode cable(s)<br />
are not the originals.<br />
Pump power is too low.<br />
Possible oxygen or<br />
'Ontamination<br />
from the purge source.<br />
Corrective Action<br />
Use the fine PHASE adjust on the Model 3955 electronics module to set the<br />
system for short, high-power pulses.<br />
Refer to "Coarse Phase Adjustment ..." in chapter 6.<br />
Increase pump power.<br />
Use only dry, oil-free, Electronic Grade 5 nitrogen gas from a cylinder. Avoid<br />
boil-off tanks which might contain oxygen or water vapor. Check the color of<br />
the desiccants in the molecular-sieve assembly in the Model 3910 and make<br />
sure they are not pink. If they are, replace the assembly (refer to chapter 9).<br />
Symptom: Satellite pulses are evident when viewed on an autocorrelator<br />
Possible Causes<br />
Room lights are too<br />
bright.<br />
Pump power is too high<br />
(ps configuration)<br />
Pump power is too high<br />
(fs configuration)<br />
Corrective Action<br />
Turn off the room lights. If the pulses disappear, they are phantom satellite<br />
pulses. To eliminate them, cover the autocorrelator to block room light.<br />
Check for real satellite pulses by turning the GTI POSITION control on the laser<br />
head control panel. If the pulses are real, i.e., they move away from or closer to<br />
the main pulse as the POSITION control is adjusted, they are caused by high<br />
pump power (cw breakthrough). Turn pump power down until the satellites disappear.<br />
Refer to appendix B, "CW Breakthrough." <strong>Tsunami</strong> power should still<br />
meet specification. If not, contact your Spectra-Physics <strong>Laser</strong>s service representative<br />
for help.<br />
If the satellite pulses are real, they disappear when you block the pump beam<br />
to the <strong>Tsunami</strong>. If they are phantom, see "Room lights" above. If you have real<br />
satellites, they are caused by high pump power (cw breakthrough). Turn pump<br />
power down until the satellites disappear. Refer to appendix B, "CW Breakthrough."<br />
<strong>Tsunami</strong> power should still meet specification. If not, contact your<br />
Spectra-Physics <strong>Laser</strong>s service representative for help.