<strong>Tsunami</strong><br />

There will be one or two reflected images visible on the card: one<br />

from M3 and perhaps one from MI. If only one image is visible,<br />

adjust M1 until two are present. The stationary image is from MS.<br />

Note that they need not over-lap at this point in the alignment.<br />

d. Remove the white card.<br />

PerJom steps e, J; h, i, and j only if you changed the focus of P2, M2 or M3<br />

during this alignment session.<br />

e. Measure along the beam path from M3 through M4 to M5, and<br />

place the white card about 45 cm from the surface of M3.<br />

f. Move the M3 translation stage until the vertically-oriented elliptical<br />

image displayed on the card becomes focused.<br />

Hold the M3 translation stage in place and loosen its setscrew (located<br />

over the dovetail slide, Figure 6-4). Carefully slide the<br />

stage until the image becomes focused on the card (Figure 6-7),<br />

then tighten the setscrew.<br />

0 C:,<br />

Mi Image M3 Image Mi and M3<br />

Images Overlapped<br />

Figure 6-7: MI and Mg Images Focused on the White Card<br />

g. Adjust M1 vertically and horizontally to overlap its reflected<br />

image onto the centered elliptical image from Ms.<br />

h. Focus M2 by moving its translation stage.<br />

Adjust the M2 micrometer to focus the weak fluorescence reflected<br />

from MI as a horizontally elliptical image on the white<br />

card.<br />

i. Remove the white card.<br />

j. Verify the pump beam is still centered on the intracavity surfaces<br />

of MI, M2. and M3.<br />

If it is not, repeat steps 12 and 13 to recenter the beam on the<br />

intracavity surfaces of MI, MP. and MS.<br />

15. Align fold mirrors M4 and Ms.<br />

a. Verify M4 and M5 are in their correct positions for ps or fs operation<br />

pigure 6-8).