<strong>Tsunami</strong><br />

Table 7-1: Bi-fi Selection for Broad Picosecond Operation<br />

Blue<br />

< 20 ps<br />

2 20 ps<br />

Standard<br />

< 20 ps<br />

2 20 ps<br />

Mid-range<br />

< 20 ps<br />

2 20 ps<br />

Long<br />

< 20 ps<br />

2 20 ps<br />

X-long<br />

< 20 ps<br />

2 20 ps<br />

Bi-Fi Type<br />

2-plate<br />

3-plate<br />

2-plate<br />

3-plate<br />

2-plate<br />

3-plate<br />

2-plate<br />

3-plate<br />

2-plate<br />

3-plate<br />

Bi-Fi Dot Color<br />

Yellow<br />

Green<br />

Yellow<br />

Green<br />

White<br />

Orange<br />

Yellow<br />

Green<br />

Yellow<br />

Green<br />

6. Verify the beam is in the center of MI0. Adjust M5 if necessary.<br />

7. Again, refer to "Optimizing <strong>Laser</strong> Output" earlier in this chapter to<br />

"walk the beam" and optimize laser output.<br />

8. Adjust the bi-fi for the desired wavelength.<br />

Part Number<br />

0434-8953<br />

0434-8973<br />

0434-8953<br />

0434-8973<br />

0434-8954<br />

0434-8974<br />

0434-8953<br />

0434-8973<br />

0434-8953<br />

0434-8973<br />

Purge the system with clean, dry nitrogen gas when you are operating<br />

within water and oxygen absorption regions (Figure 7-1).<br />

9. Push the STATUS mode locker enable button (the ENABLE LED should<br />

be lit).<br />

10. Set up an autocorrelator such as the Model 409 -08 according to its<br />

user's manual, then maximize the signal by adjusting the angle of the<br />

nonlinear crystal for the optimum phase-matching condition.<br />

11. Maximize the intensity of the autocorrelated waveform by adjusting<br />

the coarse and fine PHASE controls.<br />

Refer to chapter 6, "Alignment: Coarse Phase Adjustment ...," for<br />

infomation on setting the coarse and fine PHASE controls.<br />

12. Adjust the GTI POSITION control to obtain a pulse.<br />

13. Optimize system performance by iterating between the following<br />

adjustments:<br />

a. Adjust the fine PHASE control.<br />

b. Adjust ion laser output power to obtain a stable pulse.<br />

c. Adjust the GTI POSITION control for the desired pulse width.