<strong>Tsunami</strong><br />

b. Place a white card with a 5 mm diameter hole in it between M4<br />

and M5 SO the hole is centered on the reflected fluorescent image<br />

from M4.<br />

c. Direct the reflected image from M10 back through the hole in the<br />

card and overlap it with the image corning from M4.<br />

To determine which reflected image comes from Mlo, watch the<br />

M5 side of the card and slightly adjust Mlo until you see which<br />

image moves.<br />

d. Remove the card and lasing should begin.<br />

e. Turn the pump laser up to the optimum power found in step 8.<br />

17. Vertically and horizontally adjust Mlo and M1 for maximum output<br />

power.<br />

a. Adjust the two mirrors vertically, one at a time, for maximum<br />

output power.<br />

Iterate until maximum output power is achieved.<br />

b. Repeat, using the horizontal controls.<br />

c. Iterate steps a and b until no further increase in power is measured.<br />

d. To further optimize output, walk the cavity beam to align it to the<br />

pump beam in the rod.<br />

Refer to chapter 7, "Operation: Optimizing <strong>Laser</strong> Output."<br />

18. Set the laser output wavelength for the peak wavelength of the optics<br />

set installed. Refer to the tuning curves at the end of chapter 3. A<br />

monochromator and an ir viewer are required.<br />

a. For a fs-configured laser, install the slit and adjust it for output<br />

peak wavelength, then continue with step 19.<br />

For information on installing the slit, refer to "Converting Between<br />

fs and ps Configurations: ps to fs Conversion" later in this<br />

chapter. Remember to close the shutter during installation.<br />

b. For ps systems, install and align the bi-fi, then adjust the bi-fi for<br />

output peak wavelength. When finished, proceed with step 19.<br />

Refer to "Selecting, Installing, and Aligning the Birefringent<br />

Filter ..." later in this chapter.<br />

19. Adjust the Ti:sapphire rod for best output.<br />

a. Release the two screws on the rod translation stage that lock it in<br />

place (Figure 6-5)<br />

b. Move the rod up and down using the translation control (top<br />

knob) and position the rod near its center of travel so that a<br />

minimum amount of scatter is observed on M2.