<strong>Tsunami</strong><br />
b. Place a white card with a 5 mm diameter hole in it between M4<br />
and M5 SO the hole is centered on the reflected fluorescent image<br />
from M4.<br />
c. Direct the reflected image from M10 back through the hole in the<br />
card and overlap it with the image corning from M4.<br />
To determine which reflected image comes from Mlo, watch the<br />
M5 side of the card and slightly adjust Mlo until you see which<br />
image moves.<br />
d. Remove the card and lasing should begin.<br />
e. Turn the pump laser up to the optimum power found in step 8.<br />
17. Vertically and horizontally adjust Mlo and M1 for maximum output<br />
power.<br />
a. Adjust the two mirrors vertically, one at a time, for maximum<br />
output power.<br />
Iterate until maximum output power is achieved.<br />
b. Repeat, using the horizontal controls.<br />
c. Iterate steps a and b until no further increase in power is measured.<br />
d. To further optimize output, walk the cavity beam to align it to the<br />
pump beam in the rod.<br />
Refer to chapter 7, "Operation: Optimizing <strong>Laser</strong> Output."<br />
18. Set the laser output wavelength for the peak wavelength of the optics<br />
set installed. Refer to the tuning curves at the end of chapter 3. A<br />
monochromator and an ir viewer are required.<br />
a. For a fs-configured laser, install the slit and adjust it for output<br />
peak wavelength, then continue with step 19.<br />
For information on installing the slit, refer to "Converting Between<br />
fs and ps Configurations: ps to fs Conversion" later in this<br />
chapter. Remember to close the shutter during installation.<br />
b. For ps systems, install and align the bi-fi, then adjust the bi-fi for<br />
output peak wavelength. When finished, proceed with step 19.<br />
Refer to "Selecting, Installing, and Aligning the Birefringent<br />
Filter ..." later in this chapter.<br />
19. Adjust the Ti:sapphire rod for best output.<br />
a. Release the two screws on the rod translation stage that lock it in<br />
place (Figure 6-5)<br />
b. Move the rod up and down using the translation control (top<br />
knob) and position the rod near its center of travel so that a<br />
minimum amount of scatter is observed on M2.