Maintenance<br />
--<br />
I<br />
Prisms (fs Configuration)<br />
CAUTION<br />
DO NOT remove or adjust the prisms to clean them: they are aligned<br />
at the factory and are not to be disturbed unless specifically told to do<br />
so elsewhere in this manual. A service call might be required if disturbed.<br />
Clean them in place.<br />
1. Close the pump laser shutter.<br />
2. Clean the prisms in place. Use a separate folded tissue and a hemostat<br />
to clean the surface of each prism as outlined in step 3 of "General<br />
Procedure for Cleaning."<br />
3. Open the pump laser shutter.<br />
Tisapphire Rod<br />
<strong>Laser</strong> performance is extremely sensitive to dirt on the rod surfaces. A<br />
small portion of rod protrudes from the end of the holder to allow easy access<br />
for cleaning. Refer to Figure 6-5.<br />
WARNING<br />
NEVER remove the rod from its mount. The rotational alignment of<br />
the Ti:sapphire rod is critical and has been optimized at the factory.<br />
Any modification or adjustment other than that made by a qualified<br />
Spectra-Physics <strong>Laser</strong>s service engineer will void the warranty.<br />
1. Close the pump laser shutter.<br />
Birefringent Filter (ps configuration)<br />
2. Clean the rod surface using a folded tissue in a hemostat according to<br />
step 3 of "General Procedure for Cleaning Optics."<br />
3. Open the pump laser shutter.<br />
Typically, only the upward-facing surface of the birefringent filter gets<br />
dusty, and it can be cleaned without removing the filter.<br />
1. Close the pump laser shutter.<br />
2. Clean the top surface using a tissue and hemostat according to step 3<br />
of "General Procedure for Cleaning Optics."<br />
WARNING<br />
The birefringent filter stack is carefully aligned at the factory to orient<br />
the filter plates at Brewster's angle. NEVER disassemble the filter<br />
stack, nor remove the main birefringent filter mounting assembly from<br />
the laser.