Event Codes & Messages (Cont.)<br />

22 ANODE IS HOT E, N Anode temperature is at approximately<br />

80% of its rated heat capacity.<br />

23 N/A N/A N/A<br />

24 HOUSING<br />

OVERHEATED<br />

E, F Housing temperature is at 100% of its<br />

rated value and the thermal cutoff switch<br />

has activated open.<br />

25 WILL OVERHEAT E, N The tube anode or housing is at 80% of<br />

its rated heat capacity. Continued use in<br />

FILM mode, without allowing the tube to<br />

cool may damage the tube. You must<br />

wait for the tube to cool before the next<br />

FILM exposure is allowed, or select a<br />

sufficiently lower technique. The display<br />

will also present the message WAIT.<br />

26 N/A N/A N/A<br />

27 NO BOOST - HOT E, N The X-ray tube anode or housing is at<br />

80% of its rated heat capacity. Continued<br />

use in fluoro boost, without cooling, may<br />

damage the tube. When displayed, the<br />

BOOST ENABLE LED turns off.