14 Generator Calibration<br />

DUTY CYCLE CALIBRATION<br />

Take Duty Cycle Data<br />

NOTE: Refer to the “Calibration Tips” at the end of this procedure for things to watch for<br />

during the calibration procedure.<br />

WARNING Steps within this procedure produce X-rays. Use appropriate precautions.<br />

1. Enter <strong>com</strong>mand letter: A - Take Duty Cycle Data.<br />

A screen of instructions will appear for either the Model II or III Dynalyzer. (The<br />

model is automatically sensed by the software). Ensure that each instruction on the<br />

list has been performed before continuing.<br />

2. Depress the FOOTSWITCH (or X-RAY ON button) to select AUTO mode.<br />

3. The system will automatically acquire a series of shots (approximately 250).<br />

Observe the shots reported on the monitor. These should match those of the DRO.<br />

If not there may be a problem with the:<br />

• Auto-cal box<br />

• PIOs on the Analog Interface PCB<br />

• Cabling<br />

4. If the Dynalyzer fails to trigger, the calibration software will prompt you to press the<br />

RESET button on the DRO and then press ENTER.<br />

5. The Generator automatically switches filaments during duty cycle calibration. When<br />

this happens, the following message will appear on the Workstation screen.<br />

***SWITCHING FILAMENTS - 1 MINUTE DELAY ***<br />

NOTE: Exposures will continue until all the calibration data has been acquired.<br />

Calculate Duty Cycle Coefficients<br />

1. When the data acquisition run is <strong>com</strong>plete, you will be returned to the Model <strong>9600</strong><br />

Calibration Menu.<br />

2. Select: B - Calculate Duty Cycle Coefficients to EEPROM.<br />

3. The coefficients are calculated and briefly displayed on the screen.