Miscellaneous Error Codes<br />

One of the following codes will be displayed on the X-ray control panel display if a<br />

<strong>com</strong>munications or software error occurs. If a RT FAIL or I/O FAIL is displayed it will be<br />

in the following format: ERROR HH nnnn. HH is a hex number and nnnn is the error<br />

address.<br />

CODE DESCRIPTION<br />

FRAMING ERROR Communications error between the<br />

Control Panel Processor PCB and the<br />

Technique Processor PCB<br />

OVERFLOW ERROR Communications error between the<br />

Control Panel Processor PCB and the<br />

Technique Processor PCB<br />

PARITY ERROR Communications error between the<br />

Control Panel Processor PCB and the<br />

Technique Processor PCB<br />

RT FAIL HH nnnn Indicates a run-time software fatal error.<br />

I/O FAIL HH nnnn Indicates an I/O software fatal error.