Power Distribution 5<br />

WORKSTATION ISOLATION<br />

TRANSFORMER<br />

S/N 69-0001 thru 69-1000<br />

The isolation transformer in the Workstation provides AC to both the Workstation and<br />

the C-Arm from a single center-tapped secondary. Two phases (AC1 and AC2) of 115 V<br />

nominal, are directed to both the Workstation and the C-Arm via the Power Control PCB.<br />

BROWN<br />

WHITE/BLUE<br />

WHITE/BROWN<br />

BLUE<br />

PRIMARY SECONDARY<br />

128<br />

120<br />

110<br />

104<br />

98<br />

COM<br />

128<br />

120<br />

110<br />

104<br />

98<br />

COM<br />

PRIMARY 1<br />

PRIMARY 2<br />

AC1<br />

CT<br />

AC2<br />

98V 110V 128V 98V 110V 128V<br />

COM 104V 120V COM 104V 120V<br />

PRIMARY 1<br />

PRIMARY 2<br />

Figure 4 Isolation Transformer - Earlier Version<br />

The white/blue wire should be on Primary 1 COM, and the blue wire should be on<br />

Primary 2 COM. The brown and white/brown wires must be properly strapped to match<br />

the input line voltage as indicated by the following table:<br />

Transformer<br />

Tap<br />

Input Volt<br />

Range<br />

98V 104V 110V 120V 128V<br />

93.8<br />

to<br />

100.9<br />

101<br />

to<br />

106.9<br />

107<br />

to<br />

114.8<br />

114.9<br />

to<br />

123.9<br />

124<br />

to<br />

132