32 Power Distribution<br />

BATTERIES<br />

Two battery pack subassemblies, wired in series, which consist of sixteen 6 V batteries,<br />

wired in series, provide a nominal voltage of 192 VDC.<br />

+<br />

-<br />

BLACK<br />

BATTERY<br />

UNIT #2<br />

16 X 6 VDC=96 VDC<br />

P2B P2A<br />

BLUE BLUE<br />

192 VOLTS<br />

NOMINAL<br />

RED<br />

RED<br />

+<br />

P3A P3B<br />

RED RED<br />

-<br />

BLACK<br />

BATTERY<br />

UNIT #1<br />

16 X 6 VDC=96 VDC<br />

Battery Pack Evaluation<br />

P1B P1A<br />

BLACK BLACK<br />

Figure 19 Battery Connections<br />

CB1<br />

30A<br />

RED<br />

BLACK<br />

TO GENERATOR<br />

CONTROLLER<br />

ASSEMBLY<br />

This procedure can be used to verify whether the battery packs are performing<br />

adequately.<br />

WARNING... Steps within this procedure produce X-rays. Use appropriate precautions.<br />

1. Select FILM mode and set the technique for 75 kVp and 200 mAs.<br />

2. Make an exposure and verify that an "X-ray Overtime" message is not displayed.<br />

Verify that other technique errors such as "kV Error" and "mA Error" are not<br />

displayed. See kV Generation and mA Generation sections of this manual for<br />

details of these error messages.<br />

OR<br />

With the C-Arm covers removed, make an exposure and verify that the red LED<br />

bargraph located on the Battery Charger PCB indicates 160 to 170 volts.<br />

3. If the bargraph voltage indicates less than 160 Volts, measure across the battery<br />

terminal with a DMM while making an exposure. If the batteries measure less than<br />

160 Volts, weak battery packs or a fault with the Battery Charger PCB may be<br />

indicated.