34 Mechanical Assemblies<br />

VERTICAL COLUMN ASSEMBLY<br />

OVERVIEW<br />

Raising and lowering of the C-Arm is ac<strong>com</strong>plished by an AC motor that drives<br />

two sets of chains, a set (1 horizontal, 1 vertical) on both sides of the vertical<br />

column (see Figure 15). The 18 inch travel of the column is restricted at both<br />

ends by an upper and lower limit switch. A scale on either side of the column is<br />

provided to give an indication of the C-Arm height. Switches on top of the control<br />

panel assembly are provided for the operation of the motor.<br />

INITIALIZATION<br />

During the Mobile C-Arm boot sequence the +24 V Interlock circuit <strong>com</strong>pletes,<br />

energizing relay K3 on the Power/Motor Relay PCB. When K3 energizes 115<br />

VAC is routed to relays K12 and K13 on the Power Motor Relay PCB. The<br />

+24VDC from PS2 is also routed to relays K12 and K13 for their control by the<br />

control panel switches located on top of the cross arm housing.<br />

VERTICAL COLUMN UP<br />

When the Vertical Column lift switch is pressed the +24VDC <strong>com</strong>pletes a circuit<br />

to the DC return line (pin 1) and energizes relay K12. When K12 is energized<br />

115 VAC is switched (pin 5 to pin 6) to the vertical lift motor via the LIFT_115<br />

VAC_B line. At the same time, +24 VDC is removed from pin 8 of K13, the<br />

vertical column down relay to prevent damage to the motor in case both switches<br />

were accidentally pressed at the same time.<br />

VERTICAL COLUMN DOWN<br />

When the Vertical Column down switch is pressed the +24VDC <strong>com</strong>pletes a<br />

circuit to the DC return line (pin 1) and energizes relay K13. When K13 is<br />

energized 115 VAC is switched (pin 5 to pin 6) to the vertical lift motor via the<br />

LIFT_115 VAC_A line. At the same time, +24 VDC is removed from pin 8 of K12<br />

the Vertical Column lift relay to prevent damage to the motor in case both<br />

switches were accidentally pressed at the same time.<br />

LIMIT SWITCHES<br />

Two limit switches (LS1 and LS2) are used; one engages near full extension of<br />

the Vertical Column and one engages near full retraction. When opened the limit<br />

switches remove electrical power from the lift motor and stop the vertical column<br />

motion.