20 Generator Calibration<br />

5R THEORY - S/N 69-0001 TO 69-1160<br />

The purpose of this procedure is to ensure <strong>com</strong>pliance with FDA requirement 80-8035,<br />

Section 1020.32, part D2 for systems manufactured before May 19, 1995. This<br />

requirement refers to the maximum FLUORO dose rate output of 5R/min. To ensure<br />

<strong>com</strong>pliance, the <strong>9600</strong> is limited to 4.63 R/min in all FLUORO modes of operation (except<br />

BOOST). 4.63R/min is used instead of 5R/min to allow for worst-case equipment<br />

tolerances used to calibrate the system.<br />

To satisfy this requirement, the mA Limit File must be modified within generator<br />

software. This prevents the <strong>9600</strong> from exceeding 4.63R/min. The reason to limit the mA<br />

is shown in the following examples.<br />

Example 1<br />

With the tube current set to 5.0 mA, the 4.63 R/min is exceeded at approximately 90 kV<br />

and above. This is shown in the gray area of the graph.<br />

Figure 5 - Dose Output Exceeds 5R/Min