Generator Calibration 17<br />

VERIFICATION<br />

General Information<br />

Calibration must be verified during regular Periodic Maintenance procedures or when<br />

certified <strong>com</strong>ponents are replaced. Refer to the Appendix for a list of certified<br />

<strong>com</strong>ponents and information concerning when to file Report of Assembly, FDA Form<br />

2579.<br />

The measurement values used for these tests are not the design specifications. The test<br />

specifications used here take into account measurement limitations; their use will ensure<br />

that the design specifications are met. All reference to specifications below are to these<br />

TEST specifications.<br />

NOTE: The following tests for kVp and mAs accuracy require installation of a Dynalyzer;<br />

the Autocal Interface Unit is not required.<br />

Acquire Data for Calibration Verification<br />

WARNING... Steps within this procedure produce X-rays. Use appropriate precautions.<br />

1. From the Calibration Menu select G - Acquire Data For Calibration Verification.<br />

2. Follow the menu instructions which appear on the <strong>com</strong>puter terminal before<br />

continuing with the procedure.<br />

3. Zero the Dynalyzer (in both FLUORO and RAD on the model III). Set the Dynalyzer<br />

tank to RAD.<br />

4. Depress the FOOTSWITCH (or X-RAY ON switch) to select AUTO mode and begin<br />

acquisition run. This screen will appear on the left monitor.<br />

Change Tank Setting to RAD Mode<br />

20 Second Delay<br />

5. The system automatically acquires data for about two minutes. The software will<br />

prompt you to change the Dynalyzer III setting with the following message:<br />

*** SWITCHING FILAMENTS ***<br />

Change Tank Setting to Fluoro Mode For Small Spot Calibration.<br />

Press {ENTER} on Keyboard to Continue.<br />

6. X-rays begin after a 20 second delay. When the acquisition run is <strong>com</strong>plete<br />

(approximately 5 minutes), you will be returned to the Model <strong>9600</strong> Calibration<br />

Menu.