2 mA Generation<br />

CONTROL INPUTS<br />

The Technique Processor <strong>com</strong>mands the Analog Interface to generate four primary<br />

control signals to the X-ray Regulator PCB. Their purpose is to initialize and control the<br />

mA servo loop. These signals are:<br />

• FIL B CONTROL<br />

• MA CONTROL<br />

• FILDRVA<br />

• FILDRVB<br />

FIL B CONTROL<br />

This is a current signal from the Analog Interface PCB A/D U18-6. This current signal is<br />

converted to a positive DC voltage by the current to voltage converter U28 (pins 2 & 1).<br />

This signal is software controlled in response to the X-ray mode currently selected. It is<br />

used to set up the Filament Regulator on the Generator Driver PCB for the proper output<br />

voltage as shown in Table 1. The Filament Regulator output voltage can be measured at<br />

the collector of Q4 or at R8 on the Generator Driver PCB.<br />

Fluoro =160 VDC with X-rays off<br />

0.3 mm Film = 175 VDC, Armed 160 VDC<br />

0.6 mm Film = 185 VDC, Armed 140 VDC<br />

Table 1 - Approximate Filament Regulator Output Voltage<br />

The ratio between the FIL B CONTROL voltage and the Filament Regulator output<br />

voltage is approximately 1:40, for every +1VDC on TP10 of the Analog Interface, the<br />

filament regulator’s output will be approximately +40 VDC. See Filament Regulator<br />

Setup, later in this section, for more information.