12.01.2013 Views

9600 Service Manual - gmecorp-usa.com

9600 Service Manual - gmecorp-usa.com

9600 Service Manual - gmecorp-usa.com

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

4 Appendix<br />

SPECIFICATIONS & TOLERANCES<br />

Generator Parameters<br />

Frequency 2.5 kHz Nominal<br />

Power 4.0 kW<br />

Rectification Full Wave<br />

Modes Normal<br />

Boosted Fluoro<br />

Pulsed Fluoro<br />

kVp Accuracy ± 5% or ± 2 kVp whichever is greater<br />

mA Accuracy ± 15% or ± 0.9 mA whichever is greater<br />

mAs Accuracy ± 5% or 2 mAs<br />

kVp Rise Time Less than 10 ms<br />

kVp Overshoot Less than 5%<br />

mA Rise Time Less than 250 mS (filament limited)<br />

Ripple < 1 kVp @ 5 mA<br />

15 kVp - 20 kVp @ 100 mA<br />

X-Ray Tube Parameters<br />

Anode 2850 - 3450 rpm<br />

Focal spots 0.3 mm/0.6 mm<br />

Thermal Switch NC - opens @ 190° F<br />

High Voltage Storage 300,000<br />

Anode Cooling Rate 15,000 HU/min<br />

Housing Heat<br />

Capacity<br />

1,600,000 HU<br />

Anode Target<br />

Diameter<br />

3.1"<br />

Anode Angle 8.5°<br />

Weight 36 lbs.<br />

Pulse Mode Pulse 30 ms or 50 ms (Depends on Pulse Rate<br />

Width<br />

Selected)<br />

Linearity Film mode linearity < .08<br />

Reproducibility C.O.V. < 0.04<br />

Operating Parameters<br />

Pulse Mode Pulse 1, 2, 4, or 8 per second (selectable)<br />

Rate<br />

Pulse Mode Pulse mA 40 mA maximum<br />

kVp Range 40 - 120 kVp, fluoro mode<br />

50 - 120 kVp, film mode<br />

mA Range 0.2-5 mA normal continuous fluoro<br />

1.0 - 12 boost mode<br />

0.2-40 mA pulsed fluoro<br />

75 mA film mode<br />

mAs Range 1-300 mAs