22 Generator Calibration<br />

Example 3<br />

In the following graph, the dose rate output falls below 4 R/min at approximately 110<br />

kV. In this case, the mA limit file maximum values are too small between 110 kV to<br />

120 kV. These software mA limits would need to be increased in order to approach, but<br />

not exceed, the desired 4.63 R/min dose rate maximum limit.<br />

Figure 7 - Dose Output Incorrectly Limited