2 Image System Calibration<br />

BEAM ALIGNMENT<br />

OVERVIEW<br />

The beam alignment procedure has been designed to align imaging chain <strong>com</strong>ponents<br />

with the image intensifier. The following stages have been determined to be the best<br />

procedure for performing beam alignment in the field.<br />

1. Equipment Setup - optimizes equipment setup and positioning to minimize C-Arm<br />

flexure and the effects of the earth’s magnetic field on the image intensifier.<br />

2. CCD Camera Cooler Operation - Verify the operation of this circuit.<br />

3. Calibrate Rotation Indicator - calibrates the camera rotation before starting<br />

alignment.<br />

4. Monitor/camera Alignment Pattern - acquires the Monitor/Camera Alignment<br />

Pattern.<br />

5. Adjust Monitors - Describes how to reset size, centering and linearity of the<br />

Workstation monitors using the Monitor/Camera Alignment Pattern.<br />

6. Adjust Camera Focus - optimizes the camera focus before alignment.<br />

7. Adjust Camera Decentration -adjusts the camera’s mechanical position with respect<br />

to the output port of the image intensifier.<br />

8. Adjust Camera Electronic Centering - adjusts the camera’s electronic center with<br />

respect to the output port of the image intensifier (SW1 & SW2).<br />

9. Collimator Mechanical Center - the secondary collimator is aligned with respect to<br />

the variable collimator.<br />

10. Image Intensifier Electronic Centering - the image intensifier electronic center is<br />

located.<br />

11. Variable Collimator Centering - the entire collimator assembly is aligned with the<br />

electronic center of the image intensifier.<br />

12. Collimator Iris Calibration - the collimator iris field sizes are calibrated.<br />

13. Beam Alignment Verification - films are taken in the NORM, MAG1, MAG2 and<br />

5cm 2 positions to verify the beam alignment has been successful.