6 Power Distribution<br />

S/N 69-1001 and up<br />

The Workstation isolation transformer has two secondaries isolated from each other.<br />

One winding is a 230V center tap winding rated for 4A output. Two phases of 115V<br />

(nominal), AC1 and AC2 are directed to the Workstation via the Power Control PCB.<br />

The second winding provides 115 VAC PH and 115 VAC RTN to the C-Arm via the<br />

interconnect cable.<br />

BROWN<br />

WHITE/BLUE<br />

WHITE/BROWN<br />

BLUE<br />

PRIMARY SECONDARY<br />

105V<br />

125<br />

115<br />

105<br />

COM<br />

125<br />

115<br />

105<br />

COM<br />

COM 115V<br />

PRIMARY 1<br />

PRIMARY 2<br />

AC1<br />

CT<br />

AC2<br />

AC<br />

RTN<br />

125V 105V 125V<br />

COM<br />

115V<br />

PRIMARY 1 PRIMARY 2<br />

Figure 5 Isolation Transformer - Later Version<br />

The white/blue wire should be on Primary 1 COM, and the blue wire should be on<br />

Primary 2 COM. The brown and white/brown wires must be properly strapped to match<br />

the input line voltage as indicated by the following table:<br />

Transformer Tap 105V 115V 125V<br />

Input Volt Range 98<br />

to<br />

107<br />

108<br />

to<br />

117<br />

118<br />

to<br />

127