8 Generator Calibration<br />

11. Observing the correct polarity insert the Dynalyzer tank cable connectors into the<br />

wells of the High Voltage Tank. Be careful not to cross-thread the nuts securing the<br />

cables to the tank. Tighten with the cable spanner wrench (part number 00-877674).<br />

12. Clean the Dynalyzer wells with clean paper towels and Freon. Avoid contaminating<br />

the wells by touching the well surfaces with bare fingers.<br />

13. Clean the High Voltage cable connectors and apply silicone <strong>com</strong>pound. Observing<br />

the correct polarity install the connectors into the “Tube” side wells of the Dynalyzer<br />

tank. Be careful not to cross-thread the nuts securing the cables.<br />

NOTE: You may need to cut one or two cable-ties perform step 13.<br />

14. Plug the Dynalyzer tank, DRO and Auto-cal Interface Box into a 120 VAC outlet.<br />

Turn ON the power to all the calibration equipment and allow 30 minutes to warmup.<br />

The power outlet strip in the Workstation may be used for this purpose<br />

15. Set the mA switch on the Dynalyzer tank to Fluoro (see Figure 2) and zero the mA<br />

value (±.2mV) by adjusting the “F” mA zero trimpot. Use a DVM set to measure DC<br />

millivolts and measure at the BNC connector near the switch (see Figure 2).<br />

16. Set the mA switch on the Dynalyzer tank to Rad (see Figure 2) and zero the mA<br />

value (±.2mV) by adjusting the “R” mA zero trimpot. Use a DVM and measure at<br />

the BNC connector near the switch (see Figure 2).<br />

CAUTION: Do not use the Dynalyzer if both readings do not zero out.<br />

NOTE: The calibration software will indicate on the left monitor of the Workstation which<br />

mA switch setting (RADIOGRAPHIC/FLUORO ) on the Dynalyzer tank to use during<br />

different stages of the calibration. Set the mA switch setting as requested.<br />

17. Set the controls on the Dynalyzer III DRO (see Figure 3) as follows:<br />

Displays: A+C<br />

mA/mAs mA<br />

Trigger Source: kVp<br />

Procedure: NORMAL<br />

Trigger Mode: PRESET<br />

kV Delay: 20 ms<br />

Window Delay: OFF<br />

Trigger Level: 30<br />

Printer AUTO<br />

18. Press the RESET button on the DRO (see Figure 3). All digits should display 0.<br />

19. Turn the X-ray Switch Override on the Autocal Interface Box to ON (see Figure 4)<br />

and place the D/A -Dosimeter switch to D/A.