28 Image System Calibration<br />

IMAGE INTENSIFIER ELECTRONIC CENTER<br />

1. In the AUTO FLUORO mode, take a MAG 2 exposure. Mark the monitor at the<br />

center of the Beam Alignment Tool pattern.<br />

2. Make a NORMAL field exposure and mark the monitor at the center of the Beam<br />

Alignment Tool pattern.<br />

3. While making a NORMAL field exposure, use the target on the beam alignment tool<br />

to measure the distance between the MAG 2 and NORMAL marks. Verify the<br />

distance is less than or equal to 9 mm.<br />

NOTE: Note that each ring on the target is 1 mm wide with a 1 mm space between<br />

rings.<br />

4 Make a mark on the monitor midway between the MAG 2 and NORMAL points. The<br />

point marked is the “average electronic center”.<br />

5. Make a NORMAL field exposure. Re-adjust the Beam Alignment Mask X and Y<br />

screws until the image is centered on the average electronic center (marked in the<br />

previous step).<br />

6. Leave the Beam Alignment Mask in this position for the remainder of the Beam<br />

Alignment procedure.