Error Messages During Operation (Cont.)<br />

WARNING,<br />

HIGH KV<br />

WARNING,<br />

LOW KV<br />

WARNING,<br />

HIGH MA<br />

WARNING,<br />

LOW MA<br />

BOOST<br />

OVERTIME<br />

RELEASED<br />

EARLY<br />

kV detected higher than allowed range for desired<br />

technique.<br />

kV detected lower than allowed range for desired<br />

technique.<br />

mA detected higher than allowed range for desired<br />

technique.<br />

mA detected lower than allowed range for desired<br />

technique.<br />

The boost exposure will automatically terminate after 30<br />

seconds, even if the BOOST footswitch is still depressed.<br />

Release the BOOST footswitch <strong>com</strong>pletely and then press<br />

it again to resume boost exposures. See also System<br />

Configuration.<br />

During a film exposure, the X-RAY ON switch was<br />

released early, before the desired mAs was reached.<br />

Press the switch again and hold it down until exposure time<br />

is <strong>com</strong>plete.