Image System Calibration 33<br />

BEAM ALIGNMENT VERIFICATION<br />

1. Insert a film cassette into the Beam Alignment Tool.<br />

2. Select MANUAL FLUORO mode.<br />

3. Select the NORMAL field size and set the technique to 50 kVp @ 0.5 mA. Press<br />

the X-ray ON key for 2 seconds.<br />

4. Select MAG 1 field size and press the X-ray ON key for 2 seconds.<br />

5. Select MAG 2 field size and press the X-ray ON key for 2 seconds.<br />

6. <strong>Manual</strong>ly collimate to the minimum field size and press the X-ray ON key for 2<br />

seconds.<br />

7. Develop the film and confirm the field sizes match the following specifications:<br />

FIELD SIZE RANGE<br />

9" mode (NORMAL) A<br />

6" mode (MAG1) B<br />

4" mode (MAG2) C<br />

8. Verify that the minimum field size is less than 5 cm 2 .<br />

9. Use a ruler and verify the center of the image is centered on the sheet of film (in<br />

both the X and Y axis) within 20 mm or 13/16”.<br />

10. Verify the line pair resolution.<br />

FIELD SIZE 60 Hz lp/mm 50 Hz lp/mm<br />

NORMAL 1.4 1.2<br />

MAG1 2.0 1.7<br />

MAG2 2.4 2.0<br />

11. Send the films to OEC Medical Systems, Inc. Attention: Technical Support, 384<br />

Wright Brothers Drive, Salt Lake City, UT. 84116.<br />

12. Remove the Beam Alignment Tool.