16 Generator Calibration<br />

KV/MA TECHNIQUE CALIBRATION<br />

Acquire Calibration Data<br />

WARNING... Steps within this procedure produce X-rays. Use appropriate precautions.<br />

1. From the Model <strong>9600</strong> Calibration Menu select: D - Acquire Calibration Data.<br />

2. Remove the mA Servo-Disable jumper E1 on the X-ray Regulator PCB and change<br />

the tank setting to Radiographic mode for LARGE SPOT calibration.<br />

3. Depress the FOOTSWITCH (or X-RAY ON) to select AUTO mode and begin<br />

acquisition run.<br />

4. The system automatically acquires data for approximately three minutes and then<br />

displays the following screen:<br />

*** SWITCHING FILAMENTS ***<br />

Change tank setting to Fluoro mode for small spot calibration.<br />

Press {ENTER} on keyboard to continue.<br />

NOTE: When the *** Switching Filaments *** screen appears an audible tone will be heard<br />

until the ENTER key is pressed.<br />

5. Change the switch on the Dynalyzer to Fluoro mode and press Enter. Another<br />

series of exposures will be taken, taking about five minutes.<br />

Calculate Calibration Coefficients<br />

1. After <strong>com</strong>pletion of the acquisition run, the Model <strong>9600</strong> Calibration Menu is<br />

displayed; select E - Calculate Calibration Coefficients.<br />

2. The coefficients will be displayed briefly and then the Model <strong>9600</strong> Calibration<br />

Menu is displayed.<br />

Update Coefficients to EEPROM<br />

1. From the Calibration Menu select: F - Update Coefficients To EEPROM.<br />

2. When step one is <strong>com</strong>pleted the Model <strong>9600</strong> Calibration Menu is displayed.