Generator Calibration 33<br />

10R THEORY - S/N 69-1161 AND HIGHER<br />

The purpose of this procedure is to ensure <strong>com</strong>pliance with FDA requirement 80-8035,<br />

Section 1020.32, part D2 for systems manufactured on or after May 19, 1995. This<br />

requirement refers to the maximum FLUORO dose rate output of 10R/min.<br />

To ensure <strong>com</strong>pliance, the <strong>9600</strong> is limited to 9.26 R/min in all FLUORO modes of<br />

operation (except BOOST). 9.26R/min is used instead of 10R/min to allow for worstcase<br />

equipment tolerances used to calibrate the system.<br />

To satisfy this requirement, the mA Limit File must be modified within generator software<br />

to allow all kV stations a 5.0 mA maximum tube current.<br />

CONFIRM 10R ABS TABLES<br />

You can confirm your system has the correct 10R ABS tables by performing the<br />

following test:<br />

1. Power on the system and allow it to boot <strong>com</strong>pletely.<br />

2. Press the ABS SELECT button on the mainframe control panel. Using this key,<br />

toggle through the four available ABS tables <strong>com</strong>paring them to the following:<br />

STANDARD<br />

CHEST<br />

EXTREMITIES<br />

LOW DOSE<br />

If these options do not appear, select the correct ABS tables as described in the<br />

"10R ABS Table Configuration" procedure located on the next page.