16 Peripherals<br />

Digital Laser Camera Interface - Configuration<br />

Ensure the Workstation is configured to <strong>com</strong>municate with a Digital Laser Camera<br />

Interface PCB by performing the following steps.<br />

1. Power on the Workstation and allow it to boot <strong>com</strong>pletely.<br />

2. Insert the Diagnostic Diskette in the rear of the Workstation<br />

3. Press the SETUP OPTIONS key on the front panel as shown in Figure 13.<br />

PATIENT DISK IMAGE<br />

HELP (ANNOTATE) VIEW DIR<br />

X-RAY SPECIAL TEST<br />

SETUP<br />

MARKERS SUMMARY APPS PATTERN CUSTOMIZE OPTIONS<br />

Figure 13 - “SETUP OPTIONS” Key on Workstation Keyboard<br />

4. Select Access Level 2 and then the System Configuration option. The menu as<br />

shown in Figure 14 appears:<br />

Set configuration<br />

Line Frequency (Hz) 60<br />

System Model VAS<br />

Cine Disk ALTA4<br />

Language ENGLISH<br />

VCR1 Sony<br />

VCR2 Sony<br />

Digital Print YES<br />

Serial Number: xx-xxxx<br />

Hospital Name:<br />

Figure 14 - System Configuration Menu<br />

5. Verify the Digital Print = YES configuration entry is correct.