8 Image System<br />

COLLIMATOR<br />

OVERVIEW<br />

The iris collimator is located under a cover as shown in Figures 1 and 10. It has three<br />

motors (see Figure 5 below):<br />

• Collimator iris<br />

• Leaf rotation<br />

• Leaf in/out<br />

The operator drives a motor by pressing the desired key on the control panel. The<br />

controls shown in the left column of Table 1 are provided on both the left and right C-<br />

Arm control panels (see Figure 6).<br />

When an image field size is selected with “NORMAL,” “MAG1,” and “MAG2,” the<br />

collimator iris is opened or closed to match the selected field size. The image intensifier<br />

receives <strong>com</strong>mands at the same time and adjusts for the selected field. Information on<br />

the image intensifier is covered after the collimator section.<br />

The iris motor has a potentiometer mechanically linked to it that provides feedback for<br />

indication of iris size. The iris size may be adjusted independently of the “MAG’” size,<br />

but it cannot be adjusted to exceed the selected field size.<br />

Controls are also provided to open, close, and rotate a pair of semi-transparent copper<br />

leaves that provide further collimation of the X-ray beam.<br />