Interlocks / Stator 3<br />

GENERATOR INTERLOCK<br />

OVERVIEW<br />

The interlock circuit provides a means to halt system operation through software when the<br />

Technique Processor detects a major system fault or through hardware when a FAST<br />

STOP switch is pressed.<br />

The FAST STOP switches are located on either side of the control panel housing. The<br />

message, INTERLOCKS OPEN, will appear on both control panel displays when either<br />

FAST STOP switch is pressed. This is considered a fatal error and is recorded in the<br />

EVENT.DAT file. The system is still powered at this time and a system re-boot can be<br />

<strong>com</strong>manded by pressing any key on the control panel.<br />

The interlock circuit is <strong>com</strong>prised of the following circuits:<br />

CPU INTERLOCK<br />

• CPU INTERLOCK<br />

• FAST STOP<br />

• +24V INTERLOCK<br />

• JUMPER E5 / RELAY K4<br />

During the boot sequence the Technique Processor software toggles PIO U27-25, on the<br />

Analog Interface PCB, at a 10 mS rate. Eighteen arrows are displayed on the Control<br />

Panel display at the point in the boot sequence that this occurs.<br />

This 10 mS interlock pulse is detected by pulse detector Q1, Q2 and associated circuitry<br />

on the Analog Interface PCB. As long as the interlock pulses are present, the pulse<br />

detector holds relay K1 energized. This allows +15 VDC to pass through relay K1<br />

contacts 1 & 7 illuminating LED DS1 on the Analog Interface PCB. DS1 indicates that the<br />

signal CPU INTERLOCK (+15V) is present.<br />

The CPU INTERLOCK signal passes through the Motherboard to the Power Motor Relay<br />

PCB where it can be measured on TP3. CPU INTERLOCK momentarily energizes relay<br />

K8 on the Power Motor Relay PCB through the C7/R17 circuit. The CPU INTERLOCK<br />

signal also passes through to the FAST STOP circuit described next.