Image System Calibration 7<br />

CALIBRATE ROTATION INDICATOR<br />

WARNING Steps within this procedure produce X-rays. Use appropriate precautions.<br />

1. Boot up the system.<br />

2. Press the SETUP OPTIONS key on the Workstation keyboard. The Setup Options<br />

menu will appear.<br />

3. Insert the OEC Boot and Diagnostic disk into the disk drive.<br />

4. Select the Access Level 2 (<strong>Service</strong>) menu and press ENTER.<br />

5. Enter the Video Calibration sub-menu and run the Camera Rotation Calibration<br />

program.<br />

NOTE: This program will automatically calibrate the camera rotation indicator<br />

displayed on the left monitor to the actual degree of rotation from the stop<br />

position.<br />

6. Press ESC three times to clear the right monitor.<br />

7. During an AUTO FLUORO exposure, rotate the camera to confirm the image can<br />

rotate 360°, see Figure 4, and image movement stops when the icon stops at either<br />

limit.<br />

VIEW OF<br />

LEFT MONITOR<br />

Orientation Icon<br />

360°<br />

Image<br />

Rotation<br />

Stop Limit<br />

Figure 4 - Image Rotation Indicators