Generator Calibration 21<br />

Example 2<br />

When the mA Limit File Editor program has the correct data entered, the output of the<br />

system will resemble that of the following graph. Note the 4.63 R/min is not exceeded at<br />

90 kV and above as shown below. This is due to the software limiting the maximum<br />

available mA in all FLUORO modes as the kV for the generator increases.<br />

Figure 6 - Dose Output Correctly Limited