6 Image System Calibration<br />

VERIFY CCD CAMERA COOLER OPERATION<br />

1. With the system powered off, connect a DVM to R4 of the Image Functions PCB<br />

while observing the polarity as shown in Figure 3.<br />

R4<br />

IMAGE FUNCTION PCB<br />

DVM<br />

0.0 VDC<br />

NEG POS<br />

Figure 3 - CCD Cooler Measurement<br />

2. Power on the system and verify the CCD cooler control voltage works as follows:<br />

• 7-10 Seconds ≈ 2.5 VDC (Cooler ON)<br />

• 30-40 Seconds ≈ 0.0 VDC (Cooler OFF)<br />

3. Confirm the voltage changes for one or two cycles to verify the operation of the<br />

CCD cooler logic. Remove the DVM when finished.