18 Generator Calibration<br />

MERGE EXISTING DATA FILES<br />

1. From the Calibration Menu, select I - Return To Main Menu And Merge Existing<br />

Data Files.<br />

NOTE: Option “H” is not for service use. Skip this option.<br />

2. The message "Merging files, recreated All.Dat" will be displayed followed by a<br />

message stating that the "EEPROM. Dat file has been updated". The software then<br />

returns you to the Mainframe Menu.<br />

BACKUP GENERATOR SOFTWARE<br />

NOTE: Do not reboot the system from the Mobile C-Arm Menu or cycle power to reboot the<br />

system until the generator software has been backed up. Rebooting the system at<br />

this time affects the “Hot Byte” as explained in the “C-Arm Software” section of this<br />

manual.<br />

1. Remove the SRAM card from the Technique Processor PCB and power off the<br />

system.<br />

2. Locate the generator software backup disk (3 1/2-inch HD floppy located on the C-<br />

Arm in an anti-static bag).<br />

3. Make two 3 1/2-inch floppy backup copies of the C-Arm disk per the procedure in<br />

the C-Arm Software section of this manual. Return one backup diskette to the C-<br />

Arm and return the other to the Technical Support Office for evaluation and<br />

archiving.<br />

4. Reinstall the SRAM card in the Technique Processor and remove the cables from<br />

the Analog Interface PCB.<br />

5. Power on the system. Confirm operation.