Low Battery Charge Messages<br />

Under most circumstances the battery condition is never a problem. It is possible<br />

however, that during periods of heavy use some consideration must be made to allow<br />

adequate recharging time. The messages listed below will be displayed on the X-ray<br />

control panel fluorescent display if a battery condition error is sensed.<br />

MESSAGE DESCRIPTION<br />

70 PCT<br />

CHARGE<br />

50 PCT<br />

CHARGE<br />

When the batteries are reduced to 70 percent of their<br />

effective charge a warning message will appear on the<br />

control panel display.<br />

The message "70 PCT CHARGE" will alternate with the<br />

displayed technique.<br />

If the effective charge drops to 50 percent, the display will<br />

alternate the message "50 PCT CHARGE." Film exposures<br />

are disabled.<br />

See also Technique Processor PCB and Analog Interface PCB.<br />

Error Messages During Operation<br />

Error messages will be displayed on the X-ray Control Panel when conditions occur which<br />

result in automatic system shutdown or require system servicing.<br />

MESSAGE DESCRIPTION<br />

BATTERY<br />

DISCONNECT<br />

ED<br />

CHARGER<br />

FAILED<br />

FOOTSWITCH<br />

STUCK<br />

PRECHARGE<br />

FAIL<br />

REGULATOR<br />

FAIL<br />

BAD<br />

THERMISTOR<br />

DETECTED ...<br />

PRESS ANY<br />

KEY<br />

This error message will appear if the Battery Packs in the<br />

Generator are not connected when the system is powered up.<br />

Connect all Battery Packs and reboot the Generator.<br />

The Battery Charger circuit has failed. If this message<br />

appears during startup, the system cannot be operated. If the<br />

message appears during normal use, the system can still be<br />

operated but the Batteries will not be recharged between<br />

exposures. This message alternates with the technique<br />

display between exposures.<br />

A fault has been detected in the footswitch, X-RAY ON switch,<br />

or associated switch sensing electronics. Results in automatic<br />

shutdown. Occurs only during boot-up.<br />

The precharge cycle has failed which may indicate that the<br />

Batteries are weak. Results in automatic shutdown.<br />

A filament regulator circuit failure has been detected. The<br />

filament supply voltage is outside the desired voltage.<br />

A fault has been detected within the (X-ray Tube) housing<br />

temperature thermistor or its associated circuitry or cables.<br />

Results in automatic shutdown. Occurs only during boot-up.