40 Image System Calibration<br />

5. While making an exposure, adjust switch SW1 on the Pixel/Column Filter PCB to<br />

get the best cancellation of any column shading differences. Figure 35<br />

demonstrates what characteristics column defects look like and, therefore, what this<br />

adjustment is trying to eliminate or reduce.<br />

Figure 35 - Unacceptable Column Defect Image<br />

6. While making an exposure, adjust R1 (magnitude of the cancellation signal) on the<br />

Pixel/Column Filter PCB to increase or decrease cancellation if required. This is<br />

shown in Figure 34 and can be accessed from the side of the system as indicated<br />

by the arrow.<br />

7. Remove the copper filter.