8 C-Arm Software<br />

A - Run Applications (Take X-rays)<br />

When you have finished using this generator service software and you want to return the<br />

generator to normal operation, select option A. Run Applications (Take X-rays). This<br />

option reboots the generator. Press ESC twice to clear the right workstation monitor.<br />

Remove the diagnostic disk.<br />

B - Run Applications With Parameters<br />

When this option is selected, the following menu appears. This menu allows you to run<br />

the normal application software (which allows X-rays) with one or more parameters.<br />

Once all desired parameters have been selected, the option K. (RUN) MAIN must be<br />

selected in order to run the main application program with the parameters specified.<br />

Press ESC twice to clear the right workstation monitor.<br />

*** PARAMETERS FOR MAINFRAME APPLICATIONS ***<br />

A. (RAW) Display Sensor KV and MA after shot.<br />

B. (DRO) Display Calibrated Sensor KV and MA after shot.<br />

C. (CART) Ignore cart.<br />

D. (STATOR) Ignore stator sense.<br />

E. (NOCAL) Run machine uncalibrated.<br />

F. (HARDWARE) Ignore A/D checks during boot-up.<br />

G. (CALCOLL) Calibrate collimator.<br />

H. Change Contactor Engage Voltage (enter number):<br />

I. Delete last parameter.<br />

J. (HEAT) Display heat info.<br />

K. (RUN) MAIN<br />

Return to MAINMENU<br />

ENTER COMMAND LETTER:<br />

A. (RAW) Display Sensor kV and mA After Shot.<br />

Use this option to display the kV and mA the generator sensed after a FILM or FLUORO<br />

shot. The generator sense circuits are run uncalibrated in this mode.<br />

B. (DRO) Display Calibrated Sensor kV and mA After Shot.<br />

Use this option to display the kV and mA the generator sensed after a FILM or FLUORO<br />

shot. The generator sense circuits are run calibrated in this mode.<br />

C. (CART) Ignore Cart.<br />

CAUTION: Use this option to ignore cart status/control signals that may be disabling<br />

X-rays.<br />

D. (STATOR) Ignore Stator Sense.