Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

8 C-Arm Control<br />

X-RAY SWITCH<br />

The signals X-RAY_SWITCH and SECURITY connect at P5 on the Control<br />

Panel Processor PCB and are sent to the Analog Interface PCB via the<br />

Power/Signal Interface PCB and Mainframe Motherboard. See the X-ray<br />

On/Disable section in this manual for details of these signals.<br />

X-RAY_LAMP also connects at P5 and is sourced from the Power Motor Relay<br />

PCB. This signal provides the return path for the lamp in the X-ray switch and is<br />

derived through U1 from the signal X-RAY_ON which is the enabling signal for<br />

X-ray production. X-RAY_LAMP also goes to the Workstation to drive the LED<br />

array above the monitors.<br />

On systems with serial number 69-2001 and higher, X-RAY_SWITCH,<br />

SECURITY, and X-RAY_LAMP pass from the X-ray switch, through the new<br />

Column I/O PCB directly to the Power/Signal Interface PCB. See the X-ray<br />

On/Disable section for details of these signals.