26 Image System Calibration<br />

4. Install the Bore Sight Tool as shown in Figure 25 by performing the following:<br />

A. Remove the four nuts and washers securing the secondary filter to the collimator.<br />

����<br />

�����<br />

����<br />

Figure 25 - Installing the Bore Sight Tool<br />

B. Insert the Bore Sight Tool (with the "cup" facing up) into the secondary<br />

collimator. Line up the four bolts on the collimator so they penetrate the four<br />

holes in the Bore Sight Tool.<br />

C. Secure the Bore Sight Tool by tightening the four nuts onto the bolts.<br />

5. Take a fluoro exposure and verify that the bore sight is aimed at the center of the<br />

pattern as shown in Figure 26.<br />

Figure 26 - Image Position with Bore Sight Tool