Error Messages During Operation (Cont.)<br />

MESSAGE DESCRIPTION<br />

HV INVERTER ON High voltage was detected when it should have been off,<br />

resulting in system shutdown.<br />

KV ON IN ERROR KV was sensed without an X-ray requested. This indicates<br />

a fault with the High Voltage Generator and shuts down<br />

the system.<br />

MA ON IN ERROR MA was sensed without an X-ray requested. This indicates<br />

a fault with the High Voltage Generator and shuts down<br />

the system.<br />

STATOR NOT ON The stator is not rotating. Operation of the system without<br />

the stator rotating would be destructive to the Tube.<br />

OVERLOAD<br />

FAULT<br />

SATURATION<br />

FAULT<br />

INTERLOCKS<br />

OPEN<br />

OVERVOLTAGE<br />

FAULT<br />

Results in automatic shutdown.<br />

Indicates a fault in the high voltage regulator circuit and<br />

results in automatic shutdown. See also<br />

kV Generator.<br />

Indicates a high voltage regulator fault and results in<br />

automatic shutdown. See also kV Generator.<br />

The interlock circuit has been broken. This could be due<br />

to a fault sensed by software or a direct hardware failure.<br />

Results in automatic shutdown.<br />

A high voltage regulation failure was detected. Results in<br />

automatic shutdown. (This fault occurs at about 140 kVp to<br />

protect X-ray cables, high voltage power supply, and X-ray<br />

Tube.) See Also kV Generator.