Event Codes & Messages (Cont.)<br />

57 WARNING, HIGH KV E, N kV detected higher than allowed range<br />

for desired technique.<br />

58 WARNING, LOW KV E, N kV detected lower than allowed range for<br />

desired technique.<br />

59 WAIT N/A Message only, displayed alternately with<br />

WILL OVERHEAT<br />

5A ?? PCT CHARGE E Used as a message with battery charge<br />

state, 50 or 70 percent.<br />

5B STANDARD 5R/MIN N/A Message only, ABS table display name<br />

5C LOW DOSE N/A Message only, ABS table display name<br />

5D CHEST 5R/MIN N/A Message only, ABS table display name<br />

5E EXTREMITIES 5R/MIN N/A Message only, ABS table display name<br />

5F STANDARD N/A Message only, ABS table display name<br />

60 CHEST N/A Message only, ABS table display name<br />

61 EXTREMITIES N/A Message only, ABS table display name<br />

62 N/A N/A<br />

63 CHECKSUM ERROR N/A Message only. During boot, a checksum<br />

routine is run on the MAIN software. If<br />

checksum is in error this message will be<br />

displayed.<br />

64 BOOT-UP TERMINATED N/A Message only. Displayed after<br />

CHECKSUM ERROR (see above).