mA Generation 11<br />

MA GENERATOR ERRORS<br />

NO MA LIMITS<br />

This message appears if the MALIMIT.DAT file is not found. This file must be copied to<br />

the SRAM disk. (See the SRAM Backup and Restore sections in the C-Arm Software<br />

Section of this manual for procedures.)<br />

REGULATOR FAIL<br />

By monitoring the FILSNS voltage through the Analog Interface during an exposure, the<br />

Technique Processor software detected that the filament regulator output voltage has<br />

exceeded its voltage range by ±15% (See Table 1 for voltage ranges). This is a nonfatal<br />

error, the system shuts down for two seconds after the exposure, X-rays can then<br />

be resumed.<br />

FILAMENT FAULT<br />

With an X-ray switch engaged, the Technique Processor software has detected no<br />

voltage on the MA MEASURED line indicating little or no tube current. This is a fatal<br />

error and results in a system shut down.<br />

MA ON IN ERROR<br />

With no X-ray switch engaged, the Technique Processor software detected a voltage on<br />

the MA MEASURED line indicating tube current was present but not <strong>com</strong>manded. This<br />

is a fatal error and results in a system shut down.<br />

WARNING, HIGH MA<br />

The Technique Processor software has determined that the actual mA<br />

(MA MEASURED) exceeds the selected mA by more than 20%. This is a non-fatal error,<br />

the system shuts down for two seconds after the exposure, X-rays can then be resumed.