28 C-Arm Control<br />

BIT # PIN # SIGNAL FUNCTION COMMENT<br />

PA0 4 - Output Ma_SCALE Change mA Scale Factor circuit on LO = FILM MODE (ARM)<br />

X-ray Reg. PCB for mA Sense logic HI = FLUORO MODE<br />

PA1 3 - Output FOCLSPOT Change large/small filament in FILM LO = .6 mm; HI = 0.3 mm<br />

MODE; signal HI in FLUORO<br />

FILM MODE Only; (ARM)<br />

PA2 2 --------------- not used ---------------<br />

PA3 1 - Output SHUTDOWN Shutdown Command for generator LO = Normal Operation<br />

when fatal error/fault detected<br />

HI = Generator Shutdown<br />

PA4 40 --------------- not used ---------------<br />

PA5 39 --------------- not used ---------------<br />

PA6 38 --------------- not used ---------------<br />

PA7 37 - Output FAULTRES Reset generator fault condition to<br />

resume normal operation<br />

LO = Normal Operation<br />

HI = Fault Condition Reset<br />

PB0 18 CHRG_DIS Charger Disable HI = Battery Chrgr Disabled<br />

PB1 19 - 23 --------------- not used ---------------<br />

PB6 24 -Output XRAYON X-ray On Command for X-rays LO = X-rays OK<br />

PB7 25 - Output CPUINTLK Interlock Pulse; Start of Interlocks 20 mSec squarewave; TTL<br />

PC0 14 - Input OVRLDFLT Overload Fault; Excess Primary<br />

Current (340 A) detected with L1.<br />

PC1 15 - Input SATFAULT Saturation Fault; Excess voltage<br />

drop across Darlington transistors<br />

HI = Overload Fault<br />

PC2 16 --------------- not used ---------------<br />

PC3 17 - Input FAULT Overvoltage Fault; 140 kVp or more<br />

detected within generator assy.<br />

HI = Saturation Fault<br />

HI = Overvoltage Fault<br />

PC4 13 - Input HVGENON HV Generator On; No faults/errors HI = X-rays allowed<br />

PC5 12 - Input XRAYDISBL X-ray Disable; No faults/errors LO = X-rays allowed<br />

PC6 11 - Input INTLK Interlock Complete; interlock end LO = Interlocks OK<br />

PC7 10 - Input STATSENS Stator Sense LO = Stator OK<br />

Table 3 - U27 - PIO 2 Pin Assignments