2 C-Arm Software<br />

LOADING APPLICATION SOFTWARE<br />

Application software is by entered utilizing the operating system or from a program such<br />

as STATUS. The key events which occur are:<br />

1. Technique Processor PCB EEPROM (U20) is checked.<br />

2. Default Fluoro and Film parameters are loaded and the system is configured by the<br />

OPTIONS.DAT file. Options such as Boost and 40 mA pulsed mode are initialized.<br />

3. The 40 kHz clock on the Analog Interface PCB is started.<br />

4. Hardware is again initialized as a precaution.<br />

5 The MALIMIT.DAT file and ABS tables are loaded.<br />

6. Tests are run to determine if the system is capable of functioning.<br />

7. Footswitch is checked (for "footswitch stuck" fault).<br />

8. Generator Interlock (+24V) is made.<br />

9. Capacitors C2 and C3 on Generator Controller Assembly are charged by Battery B+<br />

voltage through pre-charge circuitry.<br />

10. The stator line is pulsed.<br />

11. Check for disconnected batteries and check B+ voltage.<br />

12. The small filament is preheated.<br />

13. Generator and X-ray tube control signals are initialized.<br />

14. Heat units are <strong>com</strong>puted for the anode and housing.<br />

15. If no problems are found, operator input is accepted from the control panel.<br />

PERIODIC FUNCTIONS AND ACTIONS<br />

These are functions executed periodically by the software program, which include<br />

Control Panel Processor PCB polling and error checking. Items checked include:<br />

1. X-ray tube heating.<br />

2. Battery voltage and current.<br />

3. Hardware faults - such as overloads, saturation, and filaments, stator, collimator,<br />

HV inverter, interlocks, <strong>com</strong>munications to Image Function PCB and Control Panel<br />

Processor PCB.